infocustech medical alarm system

 

KIT-DPX-I

 

Absorbs nuisance alarmsdpx-i

0-30 second delay

Visual indicator

Dry contact input / output

Normally Open operation  

 

 

 

 

 

The KIT-DPX-I system pictured at right provides an adjustable time delay to absorb minor single occurrence alarms.

By absorbing nuisance alarms staff will not become immune to constant audible and visual alarms.

 

  

The system utilizes our model DPX-I inline time delay, model CA-10nc Nurse Call cable and model XFORM-1 wall transformer. 

    

  

Setup & Testing 

  

1. Select a location for the DPX-I within 4’ of a 110v electrical outlet. 

2. Mount the DPX-I to a clean smooth surface with 3m Dual Lock®. 

3. Insert the CA-10nc cable into the existing dedicated Nurse Call system headwall jack.

4. Use the provided wire tie to secure the Xform-1 and CA-10nc cables from accidental disconnect. 

5. Insert the monitored device's Nurse Call output cable into the DPX-I's Nurse Call In jack. If an external system is not present the attached dummy plug is inserted to silence the alarm.

6. When the external system provides a contact closure (alarm) the DPX-I will activate the headwall jack after the adjustable time delay period is reached. It will remain active as long as the contact remains closed. It will self reset when the external contact opens. 

7. This system must be tested after each reinstallation and at regular intervals 

 

The mechanical integrity of these devices must be maintained to assure proper operation. All mounting hardware, cables, connectors and plugs must be regularly inspected for wear and tear.  All devices must be securely mounted at all times. If a mounting situation arises that does not seem safe do not use the system! Contact our technical department for advice on correcting those situations.