Remote Alarm Systems 

infocustech medical alarm system

 

KIT-NC3

 

Self contained systemKIT-NC-3

82-100db audio output

Visual indicator

Dry contact input

Normally Open operation

Battery Backup   -NEW-

 

 

 

 

The KIT-NC3 system pictured at right provides a low cost stand-alone remote alarm system with battery backup. The 11.1 volt Lithium Ion battery will provide up to eight hours of remote alarm capability when 115vac line power is lost or power is disconnected from the unit.

  

The system utilizes our model PZ-M remote lamp / buzzer, model NC-3 Nurse Call interface, and model UA-11 smart charger. A 40' double-ended modular cable completes the system.

Setup & Testing 

 

1. Select a location for the NC-3 within 6’ of a 110v electrical outlet. The battery must be tested after each reinstallation and at regular intervals.

2. Mount the NC-3 to a clean smooth surface with 3m Dual Lock®. 

3. Select a location for the PZ-M remote lamp / buzzer, normally in the hallway above the door. 

4. Install the 40’ cable between the NC-3 and PZ-M being careful to dress the cable so it remains clear of obstructions and potential damage. 

5. Use the provided wire tie to secure the UA-11 and PZ-M cables from accidental disconnect. 

6. Install a normally open dry contact ¼” mono male plug ended cable.

7. When the external system provides a contact closure the PZ-M will sound as long as the contact remains closed. It will self reset when the external contact opens. To test the 11.1 volt LI battery, unplug the UA-11 from the wall and remove the dummy plug or input cable on the NC-3. The PZ-M will alarm audibly and visually. Restore AC power (observe the full or charging lamps) to the UA-11 and reconnect the external input cable or dummy plug.

8. This system must be tested after each reinstallation and at regular intervals 

 

The mechanical integrity of these devices must be maintained to assure proper operation. All mounting hardware, cables, connectors and plugs must be regularly inspected for wear and tear.  All devices must be securely mounted at all times. If a mounting situation arises that does not seem safe do not use the system! Contact our technical department for advice on correcting those situations.

For Warranty information see;  http://www.infocustech.com/warranty.htm